Abstract

This paper reports initial results for the CLARION® Multi-Strategy™ Cochlear Implant, presently under investigational study in Europe. A magnetless implantable cochlear stimulator (ICS) with an ear-mold—supported headpiece was designed in response to an increasing demand for a magnetic resonance imaging (MRI)–compatible cochlear implant. Surgical technique, accompanying magnetless headpiece, and MRI compatibility were evaluated in 11 deaf patients (ages 6 to 62 years) who were implanted with a magnetless Clarion implant. Because of the headset mechanics, the ICS was implanted closer behind the ear than a magnet-containing ICS. The ICS-MRI compatibility was investigated with 1.5- and 0.3-T MRI. Results showed that the surgery was relatively safe and easy to learn. The headset was stable and reliable. The MRI compatibility tests indicate that the ICS poses no contraindication for patients needing MRI. Overall, the results suggest that the Clarion magnetless cochlear implant is relatively safe and easy to implant, is MRI-compatible, and functions well with the ear-mold—supported headpiece.

Keywords

CLARIONHeadsetCochlear implantImplantMedicineMagnetic resonance imagingAudiologyContraindicationBiomedical engineeringRadiologySurgeryComputer sciencePathology

Affiliated Institutions

Related Publications

Publication Info

Year
1999
Type
article
Volume
108
Issue
4_suppl
Pages
22-26
Citations
22
Access
Closed

External Links

Social Impact

Social media, news, blog, policy document mentions

Citation Metrics

22
OpenAlex

Cite This

Jürgen Neuburger, Herbert Koestler, Benno Weber et al. (1999). Clinical Results of the Clarion® Magnetless Cochlear Implant. Annals of Otology Rhinology & Laryngology , 108 (4_suppl) , 22-26. https://doi.org/10.1177/00034894991080s405

Identifiers

DOI
10.1177/00034894991080s405