Abstract

Abstract Image‐to‐image fluctuation due to physiological motion is a major limitation to the accurate detection of neuronal activity with functional MRI. In this paper, a new and general technique for the estimation and compensation of the physiological effects is presented. By simultaneously monitoring the respiration and heart beat during the acquisition of imaging data, and retrospectively synchronizing the imaging data with physiological activity, physiological effects are estimated and removed. This technique does not rely on the periodicity of the respiration or the heart beat, does not affect the signal changes arising from neuronal activation, and is beneficial to images acquired with any speed. Experimental studies performed with FLASH and EPI sequences have demonstrated that the new technique is effective in reducing physiological fluctuation and improving the sensitivity of functional MRI and is generally applicable.

Keywords

SynchronizingCommunication noiseComputer scienceHeart beatFunctional imagingRespirationArtificial intelligenceNeuroscienceMedicineInternal medicineBiology

Affiliated Institutions

Related Publications

Publication Info

Year
1995
Type
article
Volume
34
Issue
2
Pages
201-212
Citations
484
Access
Closed

External Links

Social Impact

Social media, news, blog, policy document mentions

Citation Metrics

484
OpenAlex

Cite This

Xiaoping Hu, Tuong Huu Le, Todd B. Parrish et al. (1995). Retrospective estimation and correction of physiological fluctuation in functional MRI. Magnetic Resonance in Medicine , 34 (2) , 201-212. https://doi.org/10.1002/mrm.1910340211

Identifiers

DOI
10.1002/mrm.1910340211